Factor Information
Data ID 2723
Factor Age at implantation (years)
Description The age at implantation was significantly lower in patients in whom lead failure occurred, 3.7±3.6 years versus 8.2±10.0 years (p=0.02).
Biomarker NA
Classification E1 (physiological factor - age)
Association
Application risk assessment
Objective We evaluate the incidence of epicardial lead failure and try to identify risk factors in patients with congenital heart disease.
p Value 0.02
Conclusion The age at implantation was significantly lower in patients in whom lead failure occurred, 3.7±3.6 years versus 8.2±10.0 years (p=0.02).
Risk Factor unknown
CHD Type
ID 616
CHD Type isolated CHD
CHD Subtype LVOTO/ASD/VSD/AVSD/TGA/TOF/other
Reference
PMID 21567217
Year 2011
Title Failure of epicardial pacing leads in congenital heart disease: not uncommon and difficult to predict
Sample
Population children
Source Patients' medical records and lead data were reviewed.
Region Leuven, Belgium
Method Patients' medical records and lead data were reviewed.
Race Europe
Disease History N/A
Treatment History N/A
Group (epicardial lead) Failure(Treatment) (epicardial lead) No Failure(Control)
Number 29 169
Age Age at implantation: 3.7±3.6 years Age at implantation: 8.2±10.0 years
Gender (Male: Female) 17:12 97:72
Marker Level Age at implantation: 3.7±3.6 years Age at implantation: 8.2±10.0 years